Ordering Guide of ELCOMETER Dew Point Meter G319:
Model | Model S | Model T | |
Part Number | G319—-S | G319—-T | |
Reading Parameters – RH, Ta, Ts, Td, TΔ, Tdb, Twb1, SH1 |
■ |
■ |
|
Statistics – number of readings, standard deviation, mean, coefficient of variation, minimum, maximum |
■ |
■ |
|
Dustproof & Waterproof Gauge with Fully Sealed Sensors – equivalent to IP66 |
■ |
■ |
|
Integral Magnets – secure the gauge during logging |
■ |
■ |
|
High/Low Limits – audible, visual, red/green LED alarms can be set against any or all parameters |
■ |
■ |
|
Multilingual Menus |
■ |
■ |
|
Backlight- user selectable |
■ |
■ |
|
K-Type Connector for External Measurement |
■ |
■ |
|
Memory – with reading and statistic review | Last 10 records | 25,000 records in 999 batches | |
Manual Logging |
■ |
■ |
|
Interval Logging2 | Adjustable between 1 second and 1 hour | ||
Data Output |
|
||
USB |
■ |
||
Bluetooth® to computer, Android™ & iOSǂ devices |
■ |
||
ElcoMaster® software & USB cable |
■ |
||
Certificate | • | • | |
Temperature Range | Accuracy | Resolution | |
Gauge# |
-20 to +80°C (-4 to +176°F) |
±0.5ºC (±1ºF) | 0.1ºC (0.1ºF) |
Air Temperature (Ta) | -20 to +80°C (-4 to +176°F) | ±0.5°C (±1°F)3 | 0.1°C (0.1°F) |
Surface Temperature (Ts) | -20 to +80°C (-4 to +176°F) | ±0.5°C (±1°F) | 0.1°C (0.1°F) |
External K-Type Thermocouple (Te) |
-40 to +200°C (-40 to+392°F) |
±0.5°C (±1°F)4 | 0.1°C (0.1°F) |
Relative Humidity (RH) | 0 to 100%RH | ±3%RH5 | 0.1% |
Gauge & LCD Operating Range | -20°C to +80°C (-4°F to +176°F) | ||
Power Supply | 2 x AA 1.5V Batteries or via USB Cable | ||
Battery Life | Manual Mode: Greater than 40 hours (Backlight Off) Interval Logging: up to 400 hours (1 reading every 10 minutes) |
||
Dimensions | 180 x 75 x 35mm (7 x 3 x 1.4”) | ||
Weight | 300g (0.66lb) | ||
Packing List | Elcometer 319 Dewpoint Meter, 2 x AA batteries, wrist strap, carry case, calibration certificate, USB cable+, ElcoMaster®+ and operating instructions |
STANDARDS:
BS 7079-B4, IMO MSC.215(82), IMO MSC.244(83), ISO 8502-4, US Navy NSI 009-32, US Navy PPI 63101-000
The Standards in grey have been superseded but are still recognised in some industries.
1 Calculated Value
2 With Part Number T31920162
3 Accuracy ±0.75°C below 10°C (±1.35°F below 50°F)
4 Accuracy ±2°C (4°F) with K Type probes supplied by Elcometer. Gauge tested with voltage input
5 At 1m/s
# Do not expose the gauge to temperatures outside the gauge and LCD operating range
+ Model T only
• Certificate supplied as standard
T31920162 | Magnetic Surface Temperature Probe; -40 to +80ºC (-40 to +176ºF) | ![]() |
T9996390- | Liquid Temperature Probe; -200 to +1100ºC (-328 to +2012ºF) | ![]() |
T99921325 | USB Cable | |
T99916063 |
Wrist Strap |
|
T99923480 |
Protective Carry Case/Pouch |
Ulasan